Princeton Instruments Spectroscopy Detectors

Features and Benefits:

Resource Center: Spectrographs Configurator- | Signal to Noise Calculator | WinSpec-software | Custom coated input window
 

PIXIS with permanent vacuum, maintenance free,guaranteed for life. 

Model       Data Sheet Architecture/QE Plot Pixels Pixel size µ Read noise rms
@ 100KHz
Read noise rms
@ 2MHz
Dark current         
e-/p/s
Temp.
100

front-illumin.

1340 x 100

20 x 20

2.5 e-

12 e-

0.002

-80ºC

100B

back-illumin.

1340 x 100

20 x 20

3.5 e-

13 e-

0.0025

-80ºC

100BR

back-deep-depletion

1340 x 100

20 x 20

3.5 e-

13 e-

0.04

-80ºC

100B-EX

back-illumin. eXcelon

1340 x 100

20 x 20

3.5 e-

13 e-

0.0025

-80ºC

256E

open-electrode

1024 x 256

26 x 26

6 e-

22 e-

0.002

-75ºC

256BR

back-deep-depletion

1024 x 256

26 x 26

6 e-

22 e-

0.1

-75ºC

400B

back-illumin.

1340 x 400

20 x 20

3.5 e-

13 e-

0.005

-75ºC

400BR

back-deep-depletion

1340 x 400

20 x 20

3.5 e-

13 e-

0.065

-75ºC

400B-EX

back-illumin. eXcelon

1340 x 400

20 x 20

3.5 e-

13 e-

0.005

-75ºC

2KB

back-illumin.

2048 x 512

13.5 x 13.5

2.8 e-

13 e-

0.002

-70ºC

2KB-EX

back-illumin. eXcelon

2048 x 512

13.5 x 13.5

2.8 e-

13 e-

0.002

-70ºC

2KBUV

BUV/AR coated

2048 x 512

13.5 x 13.5

3 e-

12 e-

0.002

-70ºC

PIXIS TE COOLED

SPEC-10 LN cooling with variable temp. adjust. from -70°C to -120°C     

Model       Data Sheet Architecture/QE Plot Pixels Pixel size µ Read noise rms
@ 100KHz
Read noise rms
@ 2MHz
Dark current       
e-/p/h
Temp.
100F

front-illumin.

1340 x 100

20 x 20

3 e-

12 e-

0.3

-120ºC

100B

back-illumin.

1340 x 100

20 x 20

3.5 e-

13 e-

0.3

-120ºC

100R N.A

deep-depletion

1340 x 100

20 x 20

3 e-

15 e-

1

-120ºC

100BR

back-deep-depletion

1340 x 100

20 x 20

3.5 e-

13 e-

1

-120ºC

256E

open-electrode

1024 x 256

26 x 26

6 e-

22 e-

0.5

-120ºC

256BR

back-deep-depletion

1024 x 256

26 x 26

6 e-

22 e-

2

-120ºC

400F

front-illumin.

1340 x 400

20 x 20

3 e-

12 e-

0.3

-120ºC

400B

back-illumin.

1340 x 400

20 x 20

3.5 e-

13 e-

0.3

-120ºC

400R N.A

deep-depletion

1340 x 400

20 x 20

3 e-

15 e-

1

-120ºC

400BR

back-deep-depletion

1340 x 400

20 x 20

3.5 e-

13 e-

1

-120ºC

2KB

back-illumin.

2048 x 512

13.5 x 13.5

3.5 e-

13 e-

0.3

-120ºC

2KBUV

BUV/AR coated

2048 x 512

13.5 x 13.5

3.5 e-

13 e-

0.3

-120ºC

SPEC-10 LN COOLED

OMA V  InGaAs IR camera

Model       Data Sheet Architecture/QE Plot Pixels Pixel pitch µ Read noise rms
@ High gain
Read noise rms
@ low gain
Dark current       
e-/p/s
Temp.
512-1.7 LN

Linear PDA

512 x 1

50 x 500

500 e

5000 e-

10 ke-/p/s

-120ºC

1024-1.7 LN

Linear PDA

1024 x 1

25 x 500

650 e

6000 e-

20 ke-/p/s

-120ºC

1024-2.2 LN

Linear PDA

1024 x 1

25 x 250

520 e

7000 e-

1.0 Me-/p/s

-120ºC

OMA V
OMA V
 PI-MAX ICCD
Model/Data Sheet  MCP MCP Model/QE Plot  MCP-mm Pixels Min. gate-ns 500 ps option Max. Resolution Peak QE %
PI-MAX2 1003

GEN 2

RB- | SB-| UV

18

1024 x 1024 2

available

64 lp/mm

25

PI-MAX2 1003

GEN 3 fl

HB f-| HQ f-| UNIGEN

18

1024 x 1024

2

not available

64 lp/mm

50

PI-MAX 1K

GEN 2

RB- | SB-| UV

18

1024 x 1024 2

available

64 lp/mm

45

PI-MAX 1K

GEN 3 fl

HB f-| HQ f-| UNIGEN

18

1024 x 1024 2

not available

64 lp/mm

25

PI-MAX 1024

GEN 2

RB- | SB-| UV

18

1024 x 256 2

available

64 lp/mm

45

PI-MAX 1024

GEN 3 fl

HB f-| HQ f-| UNIGEN

18

1024 x 256 2

not available

64 lp/mm

25

Sub-nano second gating: Variable gate width DC to 5 nsec and one user defined setting between 2 nsec to 500 psecUNIGEN II™ : GEN III filmless intensifier with supplementary UV response to 185 nm.
f : Gen III filmless – generation III intensifier without the ion barrier (HQ f, HB f)
HQf : highest peak quantum efficiency
HBf :high quantum efficiency with enhanced blue response
RB* : fast gate width 2 nsec, low-EBI tube for CW experiments, red/blue-balanced.
RB Slow: tube with slow gate option. Unusually high QE by eliminating the conductive underlay used for fast gating.
SB* : fast gate width 2 nsec, the super-blue photocathode response from the UV out to 500 nm.
SB Slow: tube with slow gate option. Unusually high QE by eliminating the conductive underlay used for fast gating intensifiers.
UV* : for deep UV with MgF² window for optimum sensitivity down to 120 nm.
Phosphor decay time down to 1%: P20 60 msec, P47 0.4µsec,   P46 2 µsec, P43 3 msec
EBI: Equivalent Background Illumination. Thermally generated electrons from the photocathode measured in lumens per cm²

PI-MAX